Skip to main content
Home › Guides › Gas detection › Bump test vs calibration
Guide · Gas detection · OSHA SHIB and ISEA statement

Bump test vs calibration: what each does and how often

A bump test proves the monitor still works: you hit the sensors with a known gas and watch for the alarms. It takes under a minute and does not measure accuracy. A calibration check does the same thing but compares the reading to the number on the cylinder; it must land inside the manufacturer’s tolerance, usually 10 to 20 percent. A full calibration adjusts the sensor so its reading matches the cylinder exactly. OSHA’s bulletin and the ISEA statement say to bump test or calibration check before each day’s use. A full calibration is done on the interval in the manufacturer’s manual and immediately whenever a bump test or check fails. A monitor that fails calibration comes out of service.

Bayshore Safety. Industrial safety and PPE supplier since 2017. · Current as of October 2026 · OSHA Safety and Health Information Bulletin, Calibrating and Testing Direct-Reading Portable Gas Monitors (current version, which replaced SHIB 05-04-2004); ISEA Statement on Validation of Operation for Direct Reading Portable Gas Monitors; manufacturer manuals
MSA 10045035 Calibration Gas & Kits, 20PPM Hydrogen Sulfide 60PPMMSA 10045035 Calibration Gas & Kits, 20PPM Hydrogen Sulfide 60PPMFour-gas calibration mix for LEL, O2, H2S and CO monitors

Why test a monitor that has not been touched since yesterday? Because a gas sensor fails quietly. An electrochemical cell dries out, a catalytic LEL bead is poisoned by a whiff of silicone or a sulfide gas, a sensor opening is blocked by dirt or water, and the display still reads zero. Zero is what a dead sensor reads too. OSHA’s bulletin lists the ways sensors lose accuracy: normal chemical degradation and electronic drift, extreme temperature and humidity, exposure to high concentrations of the target gas, silicones, hydrides, halogenated hydrocarbons and sulfides on LEL sensors, solvent vapors and corrosive gases on toxic sensors, and the shock of being dropped or jostled. The only way to know the monitor will alarm is to make it alarm.

Bump test, calibration check, full calibration: what is the difference?

Three different procedures that people lump together as calibration. OSHA’s bulletin defines each.

What is a bump test?A qualitative function check. A challenge gas is passed over the sensors at a concentration high enough to trip the alarms, and you confirm the gas reaches the sensors and every alarm (light, sound, vibration) fires. It does not measure accuracy.
What is a calibration check?A quantitative check. A known concentration of certified test gas is applied, and the reading must fall within the manufacturer’s acceptable range, typically 10 to 20 percent of the cylinder value. The monitor is zeroed first.
What is a full calibration?An adjustment. The instrument’s reading is set to match a known concentration of certified test gas. This is the only one of the three that changes the sensor’s response.
What happens on a fail?A monitor that fails a bump test or calibration check gets a full calibration before it is used. A monitor that fails full calibration is removed from service, and the sensor is replaced or serviced.
OSHA SHIB, Calibrating and Testing Direct-Reading Portable Gas Monitors: definitions of bump test, calibration check and full calibration; failed test actions; typical tolerance 10 to 20 percent.

How often should a gas detector be bump tested?

Before each day’s use. That is the ISEA position, which OSHA quotes in its bulletin: a bump test or calibration check of portable gas monitors should be conducted before each day’s use in accordance with the manufacturer’s instructions. OSHA adds that operators should do it daily or more often if needed, and should re-verify the monitor any time it has been dropped, exposed to a high concentration, or used in conditions that can knock a sensor off. A monitor used on two shifts gets tested before each.

OSHA SHIB, Calibrating and Testing Direct-Reading Portable Gas Monitors (ISEA statement quoted; daily or more frequently; validate operability after exposures); Draeger, bump testing guidance: visual inspection and function testing on each working day or prior to each shift.

The bump test is not written into a numbered OSHA regulation. It comes in through the confined-space rule, which requires a calibrated direct-reading instrument, and through the general duty to provide working equipment. The bulletin is OSHA’s statement of how to meet that. In a citation or after an incident, the question is whether the monitor was shown to work that day, and the bump test record is the answer.

29 CFR 1910.146(c)(5)(ii)(C) and (d)(5) (calibrated direct-reading instrument); OSHA SHIB (advisory, employers should keep calibration records for the life of each instrument).

How often should a gas detector be calibrated?

On the interval in the manual for that model. OSHA’s bulletin sets no number for full calibration; it says to follow the manufacturer’s guidelines, and to calibrate at once when a bump test or calibration check fails. Manufacturers publish intervals for each instrument, and they differ by brand, by model and by sensor; the manual, not a rule of thumb, is the authority. The daily bump test is what catches a sensor that drifts between calibrations.

OSHA SHIB, Calibrating and Testing Direct-Reading Portable Gas Monitors (follow the manufacturer’s guidelines for proper calibration; full calibration after a failed check); manufacturer manuals for the model in use.

Maintenance-free single-gas clips are the exception. Models such as the Honeywell BW Clip are sold for a fixed life of 2 or 3 years with no calibration, no sensor change and no charging; they are still bump tested before each day’s use, and the dock logs it. A monitor that logs its own bump tests, or a docking station that runs them automatically and keeps the record, is how most crews make the daily test happen.

Honeywell BW Clip published product data (2-year and 3-year maintenance-free operation; logs recent gas events and bump tests); docking station product data (MSA GALAXY GX2, Honeywell IntelliDoX, Industrial Scientific DSX).

What calibration gas do you use?

  • Certified, traceable gas, before its expiration date. OSHA’s bulletin: only use a certified traceable test gas, certified against a standard traceable to the National Institute of Standards and Technology, and never after its expiration date. Check the certificate of analysis for the cylinder.
  • The mix the manual names. A four-gas monitor is tested with a four-gas mix (a combustible gas for LEL, a reduced oxygen level, H2S and CO in a nitrogen balance); a single-gas H2S monitor with H2S in nitrogen. The LEL gas matters: a monitor calibrated on methane reads other gases differently.
  • Reactive gases have short lives. H2S, chlorine and ammonia cylinders lose concentration over time, so their shelf life is shorter than inert mixes; the expiration date on the cylinder is the limit.
  • The right regulator. Fixed-flow regulators (0.25, 0.5 and 1.0 liters per minute are the common ratings) feed a calibration cap on a diffusion monitor; demand-flow regulators are for pumped monitors, which draw what they need.
  • The right adapter. Gas is applied through the calibration cap or tubing the manufacturer specifies, so the sensors see the full concentration rather than a leak-diluted one.
OSHA SHIB, Calibrating and Testing Direct-Reading Portable Gas Monitors (certified traceable gas, expiration, certificate of analysis, reactive gases); manufacturer calibration instructions for the model (gas mix, flow rate, cap).

What does a bump test look like in practice?

  1. Turn the monitor on in clean air and let it zero. If it does not read zero (20.9 percent on oxygen) in clean air, that is the first fail.
  2. Fit the calibration cap and connect the regulator and gas. Use the cylinder and flow rate the manual names.
  3. Open the gas and watch each channel. Every sensor should rise and every alarm should fire: lights, horn and vibration. For a bump test, that is a pass. For a calibration check, each reading must also settle inside the tolerance.
  4. Close the gas, remove the cap, let the readings return to zero. A sensor that is slow to recover is a sensor to watch.
  5. Record it. Date, instrument serial number, gas lot, result. A dock does this automatically. OSHA’s bulletin says to keep calibration records for the life of each instrument.
  6. On any fail, full calibration before use. Two failed calibrations means the sensor is done: replace it or send the unit for service.
OSHA SHIB, Calibrating and Testing Direct-Reading Portable Gas Monitors (zero first, alarms activated, tolerance, records, failed-test actions); MSA, bump test and calibration guidance (two failed calibrations, remove from use); manufacturer manual for the model.

Calibration gas, regulators and docking stations

36 items · prices live

Gas mixes and concentrations are as published by the manufacturer on each product page; match the cylinder to the monitor’s manual. Free shipping on orders over $199.

ItemItemPart numberPrice
RKI Instruments 81-GX3RHSCO-LV Calibration Gas & Kits, 34 L Cylinder
ItemCalibration gasPart81-GX3RHSCO-LV
$310.25
BW Technologies CG-Q34-4 Calibration Gas & Kits, Honeywell BW 25PPM
ItemCalibration gasPartCG-Q34-4
$289.86
MSA 10048280 Calibration Gas & Kits, 34 Liter Econo-Cal Cylinder
ItemCalibration gasPart10048280
$460.65
MSA 711078 Calibration Gas & Kits, 25PPM Ammonia Balance Air
ItemCalibration gasPart711078
$704.90
MSA 10153800 Calibration Gas & Kits, 34L Calibration Testing Gas
ItemCalibration gasPart10153800
$438.90
MSA 10153803 Calibration Gas & Kits, 34L Calibration Testing Gas
ItemCalibration gasPart10153803
$495.90
MSA 10045035 Calibration Gas & Kits, 20PPM Hydrogen Sulfide 60PPM
ItemCalibration gasPart10045035
$707.09
MSA 10150622 Calibration Gas & Kits, Combustible Gas
MSA 10150622 Combustible GasUse: Calibration gas
ItemCalibration gasPart10150622
$1,152.35
MSA 803498 Calibration Gas & Kits, Calibration Cylinder
ItemCalibration gasPart803498
$478.95
MSA 711082 Calibration Gas & Kits, 500 psig Cylinder
MSA 711082 500 psig CylinderUse: Calibration gas
ItemCalibration gasPart711082
$704.90

ItemItemPart numberPrice
RKI Instruments 81-PAB-LV Calibration Kit
ItemCalibration kitPart81-PAB-LV
$425.00
RKI Instruments 81-GX3RCO Calibration Kit
ItemCalibration kitPart81-GX3RCO
$293.25
RKI Instruments 81-PAC Calibration Kit
ItemCalibration kitPart81-PAC
$403.75
RKI Instruments 81-PAB Calibration Kit
ItemCalibration kitPart81-PAB
$454.75
RKI Instruments 81-GX3RHSCO Calibration Kit
ItemCalibration kitPart81-GX3RHSCO
$327.25
RKI Instruments 81-6AAX-DLV Calibration Kit
ItemCalibration kitPart81-6AAX-DLV
$603.50
RKI Instruments 81-PAA-LV Calibration Kit
ItemCalibration kitPart81-PAA-LV
$497.25
RKI Instruments 81-PPA-LV Calibration Kit
ItemCalibration kitPart81-PPA-LV
$267.75

ItemItemPart numberPrice
Industrial Scientific 18105841 Calibration Gas & Kits, Demand Flow
ItemRegulator or capPart18105841
$621.30
MSA 10220388 Calibration Gas & Kits, ALTAIR io 4 Calibration Cap
ItemRegulator or capPart10220388
$17.95
MSA 10034391 Calibration Gas & Kits, Gas Miser Demand Calibration
ItemRegulator or capPart10034391
$964.86
MSA 467895 Calibration Gas & Kits, 0.25 LPM Fixed Flow Regulator
ItemRegulator or capPart467895
$405.12
MSA 10089321 Calibration Gas & Kits, ALTAIR 4X / 5X Calibration Cap
ItemRegulator or capPart10089321
$51.24
RKI Instruments 81-1107RK Calibration Gas & Kits, Calibration Cap
ItemRegulator or capPart81-1107RK
$19.62
RKI Instruments 81-1054RK Calibration Gas & Kits, Demand Flow, 58 L
ItemRegulator or capPart81-1054RK
$340.39
RKI Instruments 81-1051RK Calibration Gas & Kits, Regulator, Gauge
ItemRegulator or capPart81-1051RK
$161.54

ItemItemPart numberPrice
BW Technologies DX-BWC4 Gas Monitor Accessories, Honeywell BW IntelliDoX Docking
ItemDocking stationPartDX-BWC4
$3,375.20
BW Technologies DX-ULTRA Gas Monitor Accessories, Honeywell BW IntelliDoX
ItemDocking stationPartDX-ULTRA
$3,375.20
BW Technologies DX-MAXXT Gas Monitor Accessories, Honeywell BW IntelliDoX
ItemDocking stationPartDX-MAXXT
$3,375.20
BW Technologies DX-SOLO Gas Monitor Accessories, Honeywell BW IntelliDoX Docking
ItemDocking stationPartDX-SOLO
$3,375.20
Industrial Scientific 18109331-131 Gas Monitor Accessories, DSXi Cloud 3-Port
ItemDocking stationPart18109331-131
$2,542.02
BW Technologies DX-MCLP Gas Monitor Accessories, Honeywell BW IntelliDoX Docking
ItemDocking stationPartDX-MCLP
$3,375.20
Industrial Scientific 18109329-261 Gas Monitor Accessories, DSX-L Local Server
ItemDocking stationPart18109329-261
$6,459.04
Industrial Scientific Gas Monitor Accessories, DSX 3-Port Docking Station
ItemDocking stationPart18109327
$2,335.20
BW Technologies DX-CLIP Gas Monitor Accessories, Honeywell BW IntelliDoX Docking
ItemDocking stationPartDX-CLIP
$3,102.30
Industrial Scientific 18109330-131 Gas Monitor Accessories, DSXi Cloud-Connected
ItemDocking stationPart18109330-131
$2,542.02

Calibration caps, hoses and replacement sensors for each monitor are in the gas detection finder.

Questions people ask

What is the difference between a bump test and calibration?

A bump test proves the sensors respond and the alarms fire; it does not measure accuracy. A calibration check compares the reading to the cylinder value. A full calibration adjusts the reading to match the cylinder.

How often should a gas detector be bump tested?

Before each day’s use, per the ISEA statement that OSHA’s bulletin adopts, and again after a drop or an exposure that could have damaged a sensor.

How often should a gas detector be calibrated?

On the interval in the manufacturer’s manual for that model, and immediately whenever a bump test or calibration check fails. OSHA sets no fixed number.

Does OSHA require bump testing?

Not by a numbered rule. The confined-space standard requires a calibrated direct-reading instrument, and OSHA’s bulletin says the way to know it is working is a bump test or calibration check before each day’s use.

What if my monitor fails a bump test?

Full calibration before it is used. If it fails calibration, take it out of service and replace the sensor or have it serviced.

Can I use expired calibration gas?

No. OSHA’s bulletin says never to use a test gas after its expiration date; reactive gases like H2S lose concentration and will calibrate the monitor low.

What gas do I use to bump test a 4-gas monitor?

The four-gas mix named in the manual, a combustible gas at a stated percent LEL with oxygen, H2S and CO in nitrogen, through the calibration cap at the stated flow.

Do maintenance-free monitors need bump testing?

Yes. A 2-year or 3-year clip needs no calibration or sensor change, but it is bump tested before each day’s use like any other monitor.

Standards and ratings on this page are as published in the documents named and by the manufacturers named. The employer’s hazard assessment and written program decide what applies at a given workplace. Verify a rating on the product and its spec sheet before use.

Need calibration gas for a fleet, or a dock?Call (925) 215-7337 or email us the monitor models and we will match the cylinders, regulators and docks. We also service and calibrate portable gas instruments. Orders over $199 ship free.
Gas detection finder
Corporate AccountsContract pricing and inventory management available.
Live Service & SupportReal people who know the products, Monday to Friday 7:30am to 4pm Pacific. 925-215-7337
We DeliverWe ship nationwide. Parcel and LTL freight to job sites and businesses across the U.S.