Skip to main content
Home › Guides › Gas detection › 4-gas monitors
Guide · Gas detection · OSHA 1910.146 and the four sensors

What does a 4-gas monitor measure, and why those four?

A 4-gas monitor measures oxygen (as a percentage of air), combustible gas (as a percentage of the lower explosive limit), hydrogen sulfide (ppm) and carbon monoxide (ppm). Those four are the atmosphere OSHA’s confined-space standard tells you to test for, in that order: oxygen first, then flammables, then toxics, and H2S and CO are the two toxic gases met most often in tanks, sewers, pits and vaults. A reading outside 19.5 to 23.5 percent oxygen, above 10 percent LEL, or above the exposure limit for either toxic gas makes the atmosphere hazardous under the rule, and the monitor’s alarms are set at those lines.

Bayshore Safety. Industrial safety and PPE supplier since 2017. · Current as of October 2026 · 29 CFR 1910.146 and 1910.1000 Tables Z-1 and Z-2 (current e-CFR); NIOSH Pocket Guide (H2S, CO); manufacturer manuals
RKI Instruments 72-RA-C Multi-Gas Detectors, GX-3R 4-Gas Monitor LEL/O2/H2S/CORKI GX-3R 4-gas monitorLEL, O2, H2S, CO; diffusion

Why not more gases, or fewer? Four is the set that covers the standard’s definition of a hazardous atmosphere with the two toxics that are actually common. Monitors with five or six sensors exist for a fifth hazard the site knows about: sulfur dioxide at a refinery, ammonia in a cold-storage plant, carbon dioxide in a brewery, a PID for solvent vapors. A single-gas monitor is for a worker whose only hazard is one gas, a CO clip for a forklift driver. The 4-gas is the general-purpose instrument because the first three hazards are present in any enclosed space, and the fourth is in most of them.

What does each sensor measure, and what are the lines?

The four readings and where the alarms sit.

Oxygen (O2)Percent by volume; normal air is 20.9. Hazardous below 19.5 (oxygen-deficient) or above 23.5 (oxygen-enriched). Low oxygen means something else displaced it or consumed it: rust, rot, purge gas, a leak.
Combustible gas (LEL)Percent of the lower explosive limit of the calibration gas, usually methane or pentane. 100 percent LEL is the concentration that can ignite; the standard calls the atmosphere hazardous at 10 percent LEL, and that is the usual alarm.
Hydrogen sulfide (H2S)Parts per million. OSHA ceiling 20 ppm, with a 50 ppm peak for no more than 10 minutes; NIOSH ceiling 10 ppm. Smells of rotten eggs at low levels and kills the sense of smell above about 100 ppm. Typical alarms 10 and 15 ppm.
Carbon monoxide (CO)Parts per million. OSHA PEL 50 ppm averaged over 8 hours; NIOSH 35 ppm with a 200 ppm ceiling. Typical alarms 35 and 200 ppm, or 25 and 50 on some models.
29 CFR 1910.146(b), definition of hazardous atmosphere (oxygen 19.5 to 23.5 percent, 10 percent LFL, toxics above the PEL); 1910.1000 Table Z-2 (H2S ceiling 20 ppm, peak 50 ppm) and Table Z-1 (CO 50 ppm); NIOSH Pocket Guide (H2S, CO); manufacturer default alarm tables (Honeywell BW, RKI, MSA).

Why does the standard say to test in that order?

Oxygen first because the combustible sensor needs oxygen to work: a catalytic LEL sensor burns the gas on a hot bead, and in an oxygen-poor space it reads low or not at all, so a flammable reading is only trustworthy once the oxygen reading is known. Flammables second because an explosive atmosphere is the hazard that kills everyone in and around the space at once. Toxics third. The instrument reads all four at the same time; the order is about which number you believe first and which decides entry.

29 CFR 1910.146(d)(5)(iii) (test order: oxygen, combustible gases and vapors, toxic gases and vapors); manufacturer sensor documentation on catalytic-bead oxygen dependence.

Diffusion or pump?

A diffusion monitor reads the air that reaches it: worn on the chest or collar, it protects the wearer. A pumped (sample-draw) monitor pulls air through a hose, so the space can be tested from outside before anyone enters, at the top, middle and bottom, because the standard requires testing before entry and gases layer by weight. Most sites need both: a pump unit for the entry test and the attendant, diffusion units on the entrants. Several models accept a slip-on pump that converts a diffusion unit for the pre-entry check.

29 CFR 1910.146(d)(5)(i) and (c)(5)(ii)(C) (test before entry with a calibrated direct-reading instrument); manufacturer product data.

What is a bump test, and when do you calibrate?

  • Bump test: expose the monitor to a known gas before each day’s use and confirm every sensor responds and alarms. It proves the sensors are alive and the alarms work; it does not adjust anything.
  • Calibration: expose it to a certified gas of known concentration and adjust the reading to match. On the manufacturer’s schedule, typically every 30 to 180 days, and whenever a bump test fails.
  • Zero in clean air before each use; a monitor zeroed in a contaminated area will read low all day.
  • The gas cylinder is a 4-gas mix matched to the sensors (for example 50 percent LEL methane, 18 percent O2, 25 ppm H2S, 100 ppm CO, the balance nitrogen); a docking station automates both tests and keeps the record.
  • ISEA’s position, adopted in most programs: bump or calibrate before each day’s use.
ISEA Statement on Validation of Operation for Direct Reading Portable Gas Monitors (2010, reaffirmed); manufacturer manuals (Honeywell BW, RKI, MSA, Draeger); 29 CFR 1910.146(d)(4)(i) (testing and monitoring equipment) and (c)(5)(ii)(C) (calibrated direct-reading instrument).

4-gas detectors, pumps, docks and calibration gas

20 items · prices live

Sensor set, sample method and battery are as published by the manufacturer on each product page. Free shipping on orders over $199.

ItemGasesPart numberPrice
BW Technologies BWC4-Y-N Multi-Gas Detectors, / Honeywell Analytics BW, 2-Year
GasesLEL; O2; H2S; COPartBWC4-Y-N
$759.66
BW Technologies MCXL-XWHM-Y-NA Multi-Gas Detectors, / Honeywell Analytics
GasesLEL; O2; H2S; COPartMCXL-XWHM-Y-NA
$636.11
MSA 10178356 Multi-Gas Detectors, ALTAIR 4XR Multi-Gas Monitor Kit, LEL
GasesLEL; O2; H2S; COPart10178356
$2,674.65
MSA 10245598 ALTAIR io 4 Connected Multi-Gas Detector, LEL
GasesLEL; O2; H2S; COPart10245598
$1,595.24
RKI Instruments 72-RA-C-56 Multi-Gas Detectors, GX-3R 4-Gas Monitor LEL/O2/H2S/CO
GasesLEL; O2; H2S; COPart72-RA-C-56
$947.75
RKI Instruments 72-RA-C Multi-Gas Detectors, GX-3R 4-Gas Monitor LEL/O2/H2S/CO
GasesLEL; O2; H2S; COPart72-RA-C
$531.25
RKI Instruments 72-RA-C-59-NGR Multi-Gas Detectors, GX-3R LEL/O2/H2S/CO Confined
GasesLEL; O2; H2S; COPart72-RA-C-59-NGR
$1,058.25
RKI Instruments 72-RA-D Multi-Gas Detectors, GX-3R 4-Gas Monitor LEL/O2/H2S/CO
GasesLEL; O2; H2S; COPart72-RA-D
$531.25

ItemGasesPart numberPrice
BW Technologies XT-XWHM-Y-NA Sample-Draw Gas Monitors, / Honeywell Analytics, Pump
GasesLEL; O2; H2S; COPartXT-XWHM-Y-NA
$1,174.64
MSA 10116928 ALTAIR 5X Multi-Gas Monitor Kit, Probe
GasesLEL; O2; H2S; COPart10116928
$4,470.80
MSA 1016544 ALTAIR 5X Multi-Gas Monitor, LEL
GasesLEL; O2; H2S; COPart1016544
$6,829.56
MSA 10116924 ALTAIR 5X Multi-Gas Monitor, LEL
GasesLEL; O2; H2S; COPart10116924
$3,576.83
MSA 10178357 ALTAIR 4XR Multi-Gas Monitor Kit, Pump Probe
GasesLEL; O2; H2S; COPart10178357
$3,313.73
MSA 10116926 ALTAIR 5X Multi-Gas Detector Kit, LEL
GasesLEL; O2; H2S; COPart10116926
$4,034.73

ItemGasesPart numberPrice
BW Technologies 135643-L3 Calibration Gas & Kits, Honeywell BW
GasesCal gasPart135643-L3
$616.55
MSA 10242913 Calibration Gas & Kits, ALTAIR io 4 Dock Cylinder
GasesCal gasPart10242913
$1,456.59
MSA 10128644 Calibration Gas & Kits, GALAXY GX2 Test Stand, 1 Valve
GasesCal gasPart10128644
$3,684.25

Five and six-gas units, single-gas clips and replacement sensors are in the gas detection finder.

Questions people ask

What are the 4 gases in a 4-gas monitor?

Oxygen, combustible gas (percent LEL), hydrogen sulfide and carbon monoxide.

Why those four gases?

They are the OSHA confined-space definition of a hazardous atmosphere: oxygen outside 19.5 to 23.5 percent, flammables above 10 percent LEL, and toxics above their limits, with H2S and CO the two toxics most often found.

What oxygen level is safe to enter?

19.5 to 23.5 percent under 1910.146. Normal air is 20.9.

What does 10 percent LEL mean?

The combustible gas is at one tenth of the concentration that could ignite. The standard treats that as hazardous; the alarm sits there to give margin.

What are the alarm set points on a 4-gas monitor?

Manufacturer defaults are commonly O2 19.5 and 23.5 percent, LEL 10 and 20 percent, H2S 10 and 15 ppm, CO 35 and 200 ppm; the employer sets them to its program.

How often does a 4-gas monitor need calibration?

On the manufacturer’s schedule, typically 30 to 180 days, and whenever a bump test fails. Bump test before each day’s use.

Do I need a pump?

To test a space from outside before entry, yes: the standard requires testing before entry at each level. Diffusion units protect the people inside.

How long do the sensors last?

Electrochemical H2S and CO sensors and catalytic LEL sensors run about two to three years; oxygen sensors one to two years on older designs, longer on lead-free types. The monitor reports a failed sensor.

Standards and ratings on this page are as published in the documents named and by the manufacturers named. The employer’s hazard assessment and written program decide what applies at a given workplace. Verify a rating on the product and its spec sheet before use.

Outfitting a confined-space crew?Call (925) 215-7337 or email us the number of entrants and we will quote monitors, pumps, docks and gas. Orders over $199 ship free.
Gas detection finder
Corporate AccountsContract pricing and inventory management available.
Live Service & SupportReal people who know the products, Monday to Friday 7:30am to 4pm Pacific. 925-215-7337
We DeliverWe ship nationwide. Parcel and LTL freight to job sites and businesses across the U.S.