Exam gloves vs surgical gloves vs chemotherapy-rated gloves: what sterile, non-sterile, ASTM D6319, D3577 and D6978 mean, and what medical professionals wear
The FDA defines the two medical gloves by what they protect. A patient examination glove is “a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner” (21 CFR 880.6250); a surgeon’s glove is “a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination” (21 CFR 878.4460). Both are Class I medical devices that FDA reviews for leak resistance, physical properties and biocompatibility, and both have been powder-free by law since the FDA banned powdered medical gloves effective January 18, 2017. The difference in the box is sterility and quality level: surgeon’s gloves are sterile, packed as a left-and-right pair, and must meet a 1.5 AQL in FDA’s water-leak sampling (21 CFR 800.20); examination gloves are usually non-sterile, ambidextrous and boxed loose, and must meet a 2.5 AQL. “Sterile vs nitrile” is not a comparison: sterile is a packaging and processing state, nitrile is a material, and a glove can be either or both. The material standards the manufacturers print are ASTM D6319 for nitrile examination gloves, D3578 for latex examination gloves, D5250 for vinyl examination gloves and D3577 for rubber surgical gloves. A chemotherapy-rated glove is a medical glove that has also been tested under ASTM D6978 for permeation by a list of chemotherapy drugs; Radians prints ASTM D6319 and D6978 on its RWG94 examination gloves, MCR Safety prints medical grade, AQL 1.5 and ASTM D6978 on NitriMed 6012, and PIP prints ASTM D6978 on its Ambi-dex Chem and Grippaz nitrile. Medical professionals wear examination gloves for everything short of surgery, which is what OSHA’s bloodborne pathogens standard requires wherever hands may touch blood or other potentially infectious materials. This site lists examination-grade, industrial-grade and chemotherapy-tested disposable gloves and finger cots; it does not list sterile surgical gloves.
MCR Safety NitriMed 6012 Disposable Gloves, Powder Free, TexturedMCR Safety lists this glove as examination grade, 6 mil, FDA medical grade; ASTM D6978 (MCR Safety data)Why the FDA regulates a glove at all. The leak-test regulation, 21 CFR 800.20, says it in its own words: the prevalence of HIV and its risk of transmission in health care made FDA “look more closely at the quality control of barrier devices, such as surgeons’ gloves and patient examination gloves (collectively known as medical gloves),” because a device labeled as a medical glove “purport[s] to be and [is] represented to be” an effective barrier against blood- and fluid-borne pathogens. So a medical glove is a device, cleared through a 510(k) premarket notification, and FDA tests lots of them for leaks with 1,000 milliliters of water: a leak is water on the outside of the glove, and a lot whose defects exceed the acceptable quality level is adulterated. FDA set 1.5 AQL as the minimum for surgeons’ gloves and 2.5 AQL for patient examination gloves, and finger cots are excluded from the sampling plan. That is the regulatory meaning of “medical grade”: cleared as a device and held to the AQL. An industrial-grade disposable glove is the same nitrile, latex or vinyl made without the device clearance; the manufacturers on this site label the two grades on the box and on the product page (Radians: examination or industrial; MCR Safety: medical grade, AQL 1.5; PIP: exam grade). The second regulatory fact is the powder ban. FDA’s final rule of December 19, 2016 banned powdered surgeon’s gloves, powdered patient examination gloves and absorbable glove powder because of airway inflammation, hypersensitivity reactions, granulomas and adhesions; 21 CFR 895.102 now lists the powdered surgeon’s glove as a banned device, and 880.6250 and 878.4460 define the non-powdered glove as one that “does not incorporate powder for purposes other than manufacturing,” with only residual powder from manufacturing. The third fact is the workplace rule: 29 CFR 1910.1030(d)(3)(ix) requires gloves wherever an employee may have hand contact with blood, other potentially infectious materials, mucous membranes or non-intact skin, and (d)(3)(iii) requires hypoallergenic gloves, glove liners or powderless gloves to be readily accessible to employees allergic to the gloves normally provided. The bloodborne pathogens guide covers that rule in full.
The questions, one by one
From the FDA definitions, the leak-test regulation and the manufacturers’ product data.
Reading the box and the product page
The words on a disposable glove box, in the order they come up.
- Examination or exam grade: cleared by FDA as a patient examination glove and made to the 2.5 AQL water-leak limit. Radians marks its RWG94 nitrile and RWG91 latex gloves examination grade; PIP marks the Ambi-dex 2550 exam grade; MCR Safety marks NitriMed and Med Tech medical grade, AQL 1.5.
- Industrial grade: the same material without the medical-device clearance, for shop, food, janitorial and painting work; Radians RWG9405, RWG9406, RWG9606 and RWG9608 and the PIP Ambi-dex vinyl gloves are marked industrial.
- AQL: the acceptable quality level of FDA’s sampling plan under 21 CFR 800.20, from ISO 2859. 1.5 for surgeons’ gloves, 2.5 for examination gloves; a lower number is a tighter limit on leaking gloves per lot. The live guide on changing disposable gloves explains why a new glove is treated as a barrier that can fail.
- ASTM D6319: nitrile examination gloves. D3578: rubber (latex) examination gloves. D5250: PVC (vinyl) examination gloves. D3577: rubber surgical gloves. D6978: permeation by chemotherapy drugs. The manufacturer prints the ones its glove was tested to.
- EN 374-5: the European test for protection against bacteria, fungi and (where marked “virus”) viruses; PIP prints it on the Ambi-dex Chem and Grippaz gloves.
- NFPA 1999: the standard for protective clothing and ensembles for emergency medical operations; MCR Safety prints it on the Med Tech 5049 11 mil latex glove.
- Powder-free: every medical glove since January 18, 2017; the box may still say it because industrial gloves can be powdered. The black nitrile guide covers the ban.
- Sizes: examination gloves are ambidextrous and lettered, S (7), M (8), L (9), XL (10), 2XL (11) on the Radians, MCR Safety and PIP pages; the disposable glove sizes guide covers measuring for them.
Exam, surgical, chemotherapy and industrial gloves side by side
| Examination glove | Surgeon’s glove | Chemotherapy-tested glove | Industrial-grade disposable | |
|---|---|---|---|---|
| Regulation | 21 CFR 880.6250; Class I device, 510(k) | 21 CFR 878.4460; Class I device, 510(k) | An exam or surgical glove also tested to ASTM D6978 | Not a medical device |
| Purpose | Prevent contamination between patient and examiner | Protect a surgical wound from contamination | Barrier against permeation by chemotherapy drugs | Barrier for shop, food, janitorial and painting work |
| Sterility | Usually non-sterile; sterile versions exist | Sterile | Either | Non-sterile |
| Hand | Ambidextrous | Left and right pair | Either | Ambidextrous |
| AQL (21 CFR 800.20) | 2.5 | 1.5 | As the base glove | Manufacturer’s own; not FDA-sampled |
| Material standard | ASTM D6319 nitrile, D3578 latex, D5250 vinyl | ASTM D3577 rubber | ASTM D6978 in addition | Often the same ASTM test, without clearance |
| Powder | Banned since Jan 18 2017 | Banned since Jan 18 2017 | Banned | May be powdered or powder-free |
| On this site | Radians RWG94 nitrile, RWG91 latex; MCR Safety NitriMed, Med Tech; PIP Ambi-dex 2550 | Not listed | Radians RWG94; MCR Safety 6012, 6062, 5049; PIP 63-502, 63-503, Grippaz | Radians RWG9405/9406/9606/9608; PIP Ambi-dex vinyl; MCR Safety NitriShield |
The ASTM designations are as the manufacturers print them; the FDA page lists leak resistance, physical properties and biocompatibility as what it reviews.
Source: 21 CFR 880.6250, 878.4460, 800.20, 895.102; FDA Medical Gloves page; the manufacturers’ product pages on this site.
Questions people ask
What is the difference between surgical gloves and examination gloves?
A surgeon’s glove protects a surgical wound and is sterile, hand-specific and held to a 1.5 AQL; an examination glove prevents contamination between patient and examiner and is usually non-sterile, ambidextrous and held to a 2.5 AQL (21 CFR 878.4460, 880.6250, 800.20).
What is the difference between sterile and nitrile gloves?
Nitrile is a material; sterile is a packaging and processing state. Examination gloves are nitrile, latex or vinyl and usually non-sterile; surgical gloves are sterile and made of latex, polyisoprene, neoprene or nitrile.
What is the difference between surgical gloves and sterile gloves?
Every surgical glove is sterile; not every sterile glove is surgical, since sterile examination gloves are made for procedures that need a sterile field without the surgical cuff and fit.
What does it mean if gloves are non-sterile?
Not sterilized after manufacture and not packaged to keep sterility; clean examination gloves for routine patient contact and procedures that do not require a sterile field.
Are nitrile gloves surgical gloves?
Only if sold and cleared as surgeon’s gloves (sterile, paired, 1.5 AQL). The nitrile gloves on this site are examination-grade and industrial-grade; none is a surgical glove.
What are chemotherapy rated gloves?
Medical gloves tested under ASTM D6978 for permeation by chemotherapy drugs. Radians RWG94, MCR Safety NitriMed 6012 and Med Tech 5049, and PIP Ambi-dex Chem and Grippaz nitrile print the standard.
What is the difference between chemo gloves and regular gloves?
The ASTM D6978 permeation test against chemotherapy drugs, in addition to the examination-glove tests; the material is the same nitrile or latex, often thicker (5 to 8 mil on the PIP and MCR Safety styles).
What is the maximum wear time for chemotherapy gloves?
The shortest breakthrough time on the glove’s ASTM D6978 report for the drugs handled; no manufacturer on this site publishes a single number, and the standard sets none.
What kind of gloves do medical professionals use?
Examination gloves for patient contact, surgical gloves in the operating room, chemotherapy-tested gloves for hazardous drugs; nitrile has largely replaced latex because of latex allergy.
What are the types of medical gloves?
FDA names three: examination gloves, surgical gloves and chemotherapy gloves. By material, nitrile (ASTM D6319), latex (D3578), vinyl (D5250) and, for surgical gloves, rubber (D3577) and the synthetic rubbers.
What is the standard size of surgical gloves?
Surgical gloves are sized by number as hand-specific pairs; examination gloves by letter, S (7) to 2XL (11), as the Radians, MCR Safety and PIP pages print. This site lists examination gloves only.
What is used to sterilize gloves?
A sterile glove is sterilized after manufacture and sealed; the method is on the manufacturer’s label. The cleanroom garment guide describes gamma irradiation for sterile garments. No glove on this site is sold sterile.
Are there latex-free surgical gloves?
Yes; surgical gloves are made in polyisoprene, neoprene and nitrile as well as latex, and FDA requires the package to identify the material. This site does not list surgical gloves.
Why use finger cots instead of gloves?
To cover one finger (a dressing, an inspection, small electronics) without a full glove. FDA treats finger cots as examination gloves under 880.6250 but exempts them from premarket notification and the leak-test sampling plan.
Can surgical or examination gloves be used in the kitchen?
An examination glove can be a food-contact glove if its manufacturer lists the material under 21 CFR 177 or marks it FDA food contact; medical clearance by itself is not the food listing. The food handling glove guide covers it.
What does the FDA say to do with medical gloves?
Wash hands before putting on sterile gloves, make sure they fit, change them if they rip or tear, wash hands after removing them, and never reuse, wash, disinfect or share them (FDA Medical Gloves page).
Related guides
Gloves for bloodborne pathogensWhat 1910.1030 requiresDisposable gloves: when must they be changed?The rules on single-use glovesWhy black nitrile gloves?What nitrile is made of, and why latex fell out of useNitrile gloves and latex allergyAnd can you react to a non-latex glove?Nitrile vs latex vs vinyl glovesWhat each material resistsAre level 3 nitrile gloves good?What “level 3”, 4 mil and 6 mil meanThe ratings, test results and standards shown here are the manufacturers' published values, reproduced as published. Nothing on this page is a recommendation by Bayshore Safety that any product is suitable for a particular task, hazard or workplace. Selecting safety products and protective equipment is the responsibility of the end user and the employer, based on their own hazard assessment and the regulations that apply to their work. Confirm every value on the manufacturer's own documentation before use.