Skip to main content
Home › Guides › Gas detection › Calibrating RKI monitors
Guide · Gas detection · RKI GX-Force and GX-3R Pro operator’s manuals · OSHA SHIB · RKI FAQ

How to calibrate an RKI GX-Force or GX-3R Pro step by step, what gas to use, how often, and does calibration gas go bad?

The RKI Instruments manuals split calibration into two steps: a fresh air adjustment (AIR CAL) in clean air with 20.9 percent oxygen, then a span adjustment (AUTO.CAL on the GX-Force, AUTO CAL on the GX-3R Pro) with a cylinder of known gas. On the GX-Force the span uses a demand flow regulator and tubing to the pump inlet; on the GX-3R Pro it uses the calibration cup pushed on with its “front” label facing the display and a 0.25 LPM fixed flow regulator. Both manuals say the same about frequency: bump test before each day’s use with a known concentration of each gas, and calibrate only if the instrument fails the bump test, citing IEC 60079-29-2; RKI’s FAQ adds that users who bump test daily can stretch the calibration cycle to 3 to 6 months, while users who do not should calibrate monthly. The gas is the four-gas mix the instrument is set for (methane, oxygen, H2S and CO in RKI’s kits), within the concentration limits the manuals list, and it does go bad: RKI gives steel cylinders of non-reactive mixtures a three-year shelf life and aluminum cylinders of reactive mixtures (H2S, SO2, NH3, Cl2, NO2) six months to two years, with a use-by date printed on every cylinder.

Bayshore Safety. Industrial safety and PPE supplier since 2017. · Current as of November 2026 · RKI Instruments GX-Force operator's manual (PDF), Chapter 5, User Mode and Calibration (bump test, AIR CAL and AUTO.CAL procedures, Table 12 calibration concentration limits, zero suppression); RKI Instruments GX-3R Pro operator's manual (PDF), Chapter 5, User Mode and Calibration (demand zero, calibration cup, AUTO CAL and CO2 ZERO CAL procedures, calibration concentration limits); RKI Instruments FAQ page (rkiinstruments.com/resources/faqs, rkiinstruments.com), "How often should I calibrate my gas monitor?" and "Do Calibration Gases Have a Shelf Life?"; RKI Instruments SDM-3R and SDM-Force calibration station pages (rkiinstruments.com); OSHA Safety and Health Information Bulletin, Calibrating and Testing Direct-Reading Portable Gas Monitors (the three procedures: bump test, calibration check, full calibration); 49 CFR Part 172 (compressed gas as a hazardous material in transport, named only); Honeywell BW product pages, data sheets and manuals (automation.honeywell.com); MSA product pages, bulletins and operating manuals (us.msasafety.com) and the MSA 2026–2027 Safety Equipment Catalog
RKI Instruments 81-GX3RHSCO-LV Calibration Kit, H2S/CO/LEL/O2, for GX-3R ProRKI Instruments 81-GX3RHSCO-LV Calibration Kit, H2S/CO/LEL/O2, for GX-3R ProFour-gas calibration kit (cylinder, regulator and case) RKI lists for the GX-3R and GX-3R Pro (RKI data)

Why the manual puts the bump test first. RKI’s two manuals open the maintenance chapter with the same sentence: bump test the instrument before each day’s use with a known concentration of each target gas, and the instrument does not need to be calibrated unless it does not pass the bump test. That is the ISEA position OSHA’s bulletin quotes, and it matches the bulletin’s three procedures: the bump test is a function check that the gas reaches the sensors and the alarms fire; the calibration check reads a known gas and compares; the full calibration adjusts the instrument to read the cylinder value. On an RKI instrument the bump test can be run in User Mode (BUMP) or by applying gas in Measuring Mode, and with auto-calibration on (the factory setting) a failed bump test rolls straight into a calibration while the gas is still flowing. The SDM-3R and SDM-Force stations do all of it automatically and keep the record. What follows is the manual’s sequence; the manual itself is the document to work from, and this page reproduces its steps, not a substitute for them.

GX-Force, from the operator’s manual

Chapter 5, User Mode and Calibration. The GX-Force is not a gas detector while in User Mode; take it to a non-hazardous location first.

1. Enter User ModeTurn the unit off. Press and hold AIR, then press and hold POWER MODE; release both at the single beep. If a password is set (factory setting is off, password 0000), enter it with AIR and POWER MODE. BUMP appears; AIR scrolls the menu, POWER MODE enters an item and saves.
2. Fresh air adjustment (AIR CAL)In clean air of normal oxygen content (20.9 percent) with no toxic or combustible gas, scroll to GAS CAL, press POWER MODE, and with AIR CAL showing press POWER MODE again to set the zero on every channel. FAIL AIR on the screen means the adjustment failed; acknowledge and find cleaner air or check the sensor.
3. Set up the gasConfirm the instrument’s cylinder values match the label on the cylinder (manual page 76) and the cylinder assignment (page 79). Fit the demand flow regulator to the 3- or 4-gas cylinder, connect non-absorbent tubing, and if the job uses the sample hose and probe fit them to the inlet with the probe halves tight.
4. Span adjustment (AUTO.CAL)In GAS CAL, scroll to AUTO.CAL and press POWER MODE; the screen shows the gases assigned to Cylinder A and their values. Make sure the unit has been on at least 45 seconds, connect the tubing to the inlet or the probe, press POWER MODE, let the gas flow for 1 minute, press POWER MODE and read the result: P or F per channel.
5. FinishDisconnect the tubing, unscrew the regulator, and press POWER MODE to START to enter Measuring Mode (with auto start after a successful calibration on, the unit does it). A channel that fails is calibrated again; a sensor that will not calibrate is replaced. MAX.SPAN, if turned on, flags a sensor whose adjustment is near its limit as due for replacement.
Concentration limits (Table 12)Combustible gas 1 to 75 percent LEL; oxygen 0.0 to 17.0 percent; hydrogen sulfide 0.5 to 200.0 ppm; carbon monoxide 12 to 2,000 ppm. The hydrogen response of a hydrogen-compensated CO sensor is calibrated monthly and when the sensor is replaced.

GX-3R Pro, from its operator’s manual

From Chapter 5 of the GX-3R Pro manual and the SDM pages:

  • Demand zero (fresh air): in clean air, the DEMAND ZERO item in Maintenance Mode or the AIR button in Measuring Mode zeroes the channels; AUTO ZERO can be set on or off.
  • Set up: install the calibration cup, pushed all the way on, using its label and the imprinting so it sits in the correct orientation with the "front" label visible from the display side; connect the tubing from a 0.25 LPM fixed flow regulator to the cup's inlet.
  • Vent the regulator while installing it on a toxic gas cylinder (H2S), turning the knob counterclockwise to open it as it goes on, to keep air from entering the cylinder and degrading the gas.
  • Span (AUTO CAL): in User Mode place the cursor on GAS CAL, press POWER MODE, place the cursor on AUTO CAL and press POWER MODE; the display shows the gases assigned to Cylinder A and their values (confirm they match the cylinder label, manual page 87). Make sure the unit has been on at least 45 seconds, press POWER MODE, open the regulator, let the gas flow for 1 minute (2 minutes for ammonia), press POWER MODE and follow the pass/fail result; close the regulator, unscrew it, remove the cup and store the kit.
  • CO2 ZERO CAL: the infrared CO2 sensor has its own zero step in the Gas Cal menu, run with the zero gas (nitrogen) from RKI's CO2 kits for 1 minute.
  • Concentration limits: combustible 5 to 75 percent LEL; oxygen 0.0 to 17.0 percent; H2S 1.0 to 200.0 ppm; CO 15 to 2,000 ppm; hydrogen 25 to 2,000 ppm; ammonia 8 to 400 ppm; CO2 3,000 to 9,000 ppm or 1.00 to 4.00 percent; HCN 0.9 to 30.0 ppm; NO2 0.50 to 20.00 ppm; PH3 0.05 to 20.00 ppm; SO2 0.50 to 100.00 ppm.
  • Or the dock: RKI's SDM-3R charges, calibrates and bump tests the GX-3R and GX-3R Pro and stores every record for USB upload or PC control of up to ten stations; the SDM-Force does the same for the GX-Force.
RKI Instruments GX-3R Pro operator's manual (PDF), Chapter 5, User Mode and Calibration; RKI Instruments SDM-3R and SDM-Force calibration station pages (rkiinstruments.com).
Calibration gas concentrations the RKI manuals accept
ChannelGX-Force minimumGX-Force maximumGX-3R Pro minimumGX-3R Pro maximum
Combustible gas1% LEL75% LEL5% LEL75% LEL
Oxygen0.0%17.0%0.0%17.0%
Hydrogen sulfide0.5 ppm200.0 ppm1.0 ppm200.0 ppm
Carbon monoxide12 ppm2,000 ppm15 ppm2,000 ppm
Hydrogen (for the H2-compensated CO sensor)——25 ppm2,000 ppm
Ammonia——8 ppm400 ppm
Carbon dioxide, ppm sensor——3,000 ppm9,000 ppm
Carbon dioxide, percent sensor——1.00% vol4.00% vol
Hydrogen cyanide——0.9 ppm30.0 ppm
Nitrogen dioxide——0.50 ppm20.00 ppm
Phosphine——0.05 ppm20.00 ppm
Sulfur dioxide——0.50 ppm100.00 ppm

Source: RKI Instruments GX-Force operator's manual (PDF), Chapter 5, User Mode and Calibration, Table 12; RKI Instruments GX-3R Pro operator's manual (PDF), Chapter 5, User Mode and Calibration, calibration concentration limits

Calibration gas shelf life, as RKI lists it
CylinderGas mixtureShelf life (RKI warranty)
17, 34 or 103 L steelAll mixtures3 years
34 or 58 L aluminumH2S in nitrogen2 years
34 or 58 L aluminumLEL / O2 / H2S / CO (four-gas)2 years
34 or 58 L aluminumSO2 in nitrogen2 years
34 or 58 L aluminumNH3 in nitrogen18 months
34 or 58 L aluminumCl2 in nitrogen9 months
34 or 58 L aluminumHCl in nitrogen1 year
34 or 58 L aluminumHCN in nitrogen18 months
34 or 58 L aluminumNO in nitrogen18 months
34 or 58 L aluminumNO2 in nitrogen6 months
34 or 58 L aluminumPH3 in nitrogen18 months
34 or 58 L aluminumSiH4 in nitrogen1 year

Source: RKI Instruments FAQ page (rkiinstruments.com/resources/faqs, rkiinstruments.com), "Do Calibration Gases Have a Shelf Life?". RKI states that every one of its cylinders is printed with a use-by month and year and should not be used past it.

Questions people ask

How to calibrate step by step?

From RKI Instruments’ GX-Force manual: take the instrument to fresh air; hold AIR then POWER MODE until one beep to enter User Mode; scroll to GAS CAL and run the fresh air adjustment (AIR CAL); scroll to AUTO.CAL and check the cylinder values on the screen against the label; fit the demand flow regulator and tubing to the inlet with the unit on at least 45 seconds; apply gas for 1 minute; press POWER MODE and read pass or fail for each channel. The GX-3R Pro uses the calibration cup and a 0.25 LPM fixed flow regulator with the same two steps.

How to calibrate an RKI GX force?

AIR CAL in clean air, then AUTO.CAL with the four-gas cylinder through a demand flow regulator to the pump inlet, as the GX-Force manual’s Chapter 5 lays out; the concentration limits it accepts are combustible gas 1 to 75 percent LEL, oxygen 0 to 17 percent, H2S 0.5 to 200 ppm and CO 12 to 2,000 ppm.

How to calibrate a H2S meter?

The same two steps on the H2S channel: zero in clean air, then span with H2S calibration gas at a concentration the manual accepts (RKI: 1.0 to 200.0 ppm on the GX-3R Pro, 0.5 to 200 ppm on the GX-Force; the kits carry 25 ppm H2S in the four-gas mix). RKI’s manual says to vent the regulator while fitting it to a toxic cylinder so air does not enter; H2S cylinders in aluminum carry a 2-year shelf life.

How often should a bump test be performed?

Before each day’s use, with a known concentration of each gas, per both RKI manuals (citing IEC 60079-29-2), OSHA’s bulletin on direct-reading monitors and the ISEA statement; calibrate only if the bump test fails.

How often does an RKI monitor need calibration?

RKI’s FAQ: users who bump test daily can extend the calibration cycle to 3 to 6 months for instruments that pass; users who do not should calibrate monthly; a user may start weekly and lengthen the interval as adjustments stay small, which for most users ends between one and three months.

What are the three types of calibration?

In the manuals’ terms: the fresh air adjustment (zero), the span adjustment with calibration gas, and the bump test that checks both without adjusting. A docking station (SDM-3R, SDM-Force) runs all three automatically and stores the records.

Does calibration gas go bad?

Yes. RKI prints a use-by month and year on every cylinder: 3 years for steel cylinders, 2 years for aluminum H2S, four-gas and SO2 cylinders, 18 months for ammonia, HCN, nitric oxide and phosphine, 1 year for HCl and silane, 9 months for chlorine, 6 months for nitrogen dioxide.

How is a multi-gas detector calibrated?

All channels at once with a mixed cylinder: the manual accepts the cylinder values within the concentration limits in the table on this page, the display shows the gases assigned, and each channel passes or fails on its own. The CO2 channel on a GX-3R Pro has a separate CO2 ZERO CAL with nitrogen because outdoor air already holds about 400 ppm.

Is calibration gas hazmat?

In transport, yes: a compressed gas cylinder ships as a hazardous material under 49 CFR by ground. In use it is released at 0.25 to 0.5 liters per minute into the calibration cup or tubing; the SDS lists the hazards of the mix.

What is a calibration certificate?

A document RKI offers with a new instrument (part 90-CALCERT, requested at order) listing the model, part and serial numbers, the pre-shipment readings and the traceability of the gas used; RKI describes itself as an ISO 9001 company in the same FAQ answer.

The ratings, test results and standards shown here are the manufacturers' published values, reproduced as published. Nothing on this page is a recommendation by Bayshore Safety that any product is suitable for a particular task, hazard or workplace. Selecting safety products and protective equipment is the responsibility of the end user and the employer, based on their own hazard assessment and the regulations that apply to their work. Confirm every value on the manufacturer's own documentation before use.

Need monitors matched to the gases on your permit?Call (925) 215-7337 or email us the gases in your hazard assessment and where people work, and we will quote the monitors, calibration gas and docks with the manufacturer figures listed. We service and calibrate portable gas instruments.
Gas detection selector
How to buyCard, purchase order or invoice terms; quotes on request; agencies and institutions welcome.
Live Service & SupportReal people who know the products, Monday to Friday 7:30am to 4pm Pacific. 925-215-7337
We DeliverWe ship nationwide. Parcel and LTL freight to job sites and businesses across the U.S.