Description
The Draeger 8103571 is the Hydrocarbons 0.1%/c detector tube, measuring range 0.1 to 1.3 Vol% propane, 1 pump strokes, measuring time max. 3 min (handbook, p. 191). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Hydrocarbons.
Draeger 8103571 specifications
| Item | Draeger data |
|---|---|
| Part number | 8103571 |
| Gas or substance | Hydrocarbons |
| Tube type | 0.1%/c |
| Measuring range | 0.1 to 1.3 Vol% propane |
| Number of strokes (n) | 1 |
| Measuring time | max. 3 min |
| Standard deviation | ± 15 % |
| Color change | orange → brown-green |
| Temperature | 0 to 40 °C |
| Humidity | 1 to 40 mg/L |
| Reaction principle | C3H8/C4H10 + Cr6+ → Cr3+ + var. oxidation products. |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 191, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 191): The information regarding the cross-sensitivity only applies for measurements with a maximum of 1 stroke. Hydrocarbons and hydrocarbons with olefin double bonds are indicated with a different discoloration and sensitivity. The following have no effect on the reading of 0.1 Vol% propane/butane: < 99.9 Vol% methane < 5 Vol% ethane < 1 Vol% carbon monoxide < 500 ppm acetylene, ethylene
Note (handbook, p. 191): A maximum of 15 strokes within 1 hour can be performed for leak- age measurements (qualitative measurements). However, the infor- mation regarding the cross-sensitivity only applies for measurements with a maximum of 1 stroke!
Other Draeger tubes for Hydrocarbons
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| 8103581 | Draeger Hydrocarbons 2/a Detector Tubes, Box of 10 | 2 to 24 mg/L | Box of 10 |
Every Draeger tube for this gas, with live prices: Hydrocarbons detector tubes.
Draeger 8103571 questions
What does the Draeger 8103571 measure? Hydrocarbons in air, measuring range 0.1 to 1.3 Vol% propane.
How many pump strokes does the 8103571 take? 1 strokes (n) per the handbook, p. 191. Measuring time max. 3 min.
Which pump does the 8103571 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the 8103571? orange → brown-green, read against the scale printed on the tube.
Do Draeger 8103571 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of 8103571? 10. Each tube is used once.
Guides: what Draeger tubes measure · pump strokes and measuring time · how to read a tube · expiration and storage · cross-sensitivity · all tubes by gas and range.