How many pump strokes does a Draeger tube take, and how long does a reading take?

The number of strokes is set per tube and per scale and is printed on the tube next to that scale; it ranges from 1 stroke (Ammonia 5/a, high range) to 100 strokes (Formaldehyde 0.2/a with the activation tube). One stroke of the Draeger accuro or X-act 5000 draws 100 mL (± 5 %), so the strokes define the air volume the scale was calibrated for, and the defined measuring range applies only with a Draeger pump (handbook pp. 30, 33, 114). The measuring time is the average time for that range: the handbook gives 10 seconds to 15 minutes as the span for short-term tubes, and every tube entry states its own, for example approx. 1.5 min for Formaldehyde 0.2/a at 10 strokes and approx. 3 min at 20 strokes (pp. 35, 181).

Bayshore Safety. Industrial safety and PPE supplier since 2017. · Current as of November 2026 · Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), section 2.3 (pp. 28–35), section 2.4 (pp. 35–36), section 6.1.1 (pp. 114–115) and the tube entries (pp. 117 onward)
Draeger 6400000 accuro detector tube hand pumpDraeger accuro detector tube hand pumpBellows pump, 100 mL (± 5 %) per stroke, 1 to 50 strokes and higher, no power supply (handbook p. 31)

Why a fixed number of strokes? The handbook describes the tube and pump as a single measuring system: each type of tube is calibrated in batches together with a Draeger pump, and the scale assumes exactly the volume those strokes draw. Draeger and the pump standards therefore require a tube to be used with the matching pump from the same manufacturer; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the function of the measurement system (pp. 28–29). On the accuro the end of a stroke is shown by a pressure-controlled end-of-stroke indicator in the pump head; on the X-act 5000 the stroke count is entered and the pump runs it automatically, 1 to 199 strokes (pp. 30, 33).

Strokes, volume and time, by the handbook

Stroke volume100 mL (± 5 %) on both the accuro and the X-act 5000 Basic (pp. 30, 33). Ten strokes draw one liter.
Strokes per scalePrinted on the tube next to each scale. Scale tubes: the strokes relate directly to the scale numbers. Color-comparison and marking-ring tubes: a lowest and highest stroke count until a color pattern appears (p. 114).
Two scales, two countsThe high range needs fewer strokes, the low range more. Start with the high-range strokes; add the difference and read the second scale if a finer reading is possible (p. 36).
Measuring timeThe average time for the range, in s or min, stated per tube (p. 115). The X-act 5000’s pump control shortens the average time for tubes with high stroke numbers compared with the hand pump (p. 30).
Range extensionWhere a changed stroke count extends the range, the entry states the extended range and the correction factor, for example Formaldehyde 0.2/a: 40 strokes, divide by 2; 80 strokes, divide by 4; 100 strokes, divide by 5, with the activation tube 8101141 (p. 181).
Before and afterA leak and suction check before every measurement, and a few empty strokes without a tube after it to purge reaction products from the bellows (p. 34).

Examples from the handbook entries (strokes / measuring time)

  • Ammonia 5/a (CH20501): 10 strokes / approx. 60 s on the 5 to 70 ppm scale; 1 stroke / approx. 10 s on the 50 to 700 ppm scale.
  • Formaldehyde 0.2/a (6733081): 10 strokes / approx. 1.5 min for 0.5 to 5 ppm; 20 strokes / approx. 3 min for 0.2 to 2.5 ppm (p. 181).
  • Formaldehyde 2/a (8101751): 5 strokes / approx. 30 s for 2 to 40 ppm (p. 182).
  • Acetaldehyde 100/a (6726665): 20 strokes for 100 to 1000 ppm.
  • Every tube page on this site prints its own strokes and measuring time from the handbook entry, under "specifications".

Tubes with low, middle and high stroke counts

6 items · prices live

Strokes and times from the handbook entries; the tube page has the full entry. Orders ship in 1–2 business days.

ItemRangePart numberPrice
Draeger Ammonia 5/a Detector Tubes, Box of 10
Draeger Ammonia 5/a Detector Tubes, Box of 10Handbook: 10 /1 strokes, approx. 60 s / approx. 10 s · Sold by the box
Range5 to 70 ppm / 50 to 700 ppmPartCH20501
$129.45
Draeger Formaldehyde 2/a Detector Tubes, Box of 10
Draeger Formaldehyde 2/a Detector Tubes, Box of 10Handbook: 5 strokes, approx. 30 s · Sold by the box
Range2 to 40 ppmPart8101751
$129.57
Draeger Acetone 40/a Detector Tubes, Box of 5
Draeger Acetone 40/a Detector Tubes, Box of 5Handbook: 1 strokes, approx. 1 min · Sold by the box
Range40 to 800 ppmPart8103381
$115.98

ItemRangePart numberPrice
Draeger Formaldehyde 0.2/a Detector Tubes, Box of 10
Draeger Formaldehyde 0.2/a Detector Tubes, Box of 10Handbook: 10 / 20 strokes, approx. 1.5 min / approx. 3 min · Sold by the box
Range0.5 to 5 ppm / 0.2 to 2.5 ppmPart6733081
$128.95
Draeger Acetaldehyde 100/a Detector Tubes, Box of 10
Draeger Acetaldehyde 100/a Detector Tubes, Box of 10Handbook: 20 strokes, approx. 5 min · Sold by the box
Range100 to 1000 ppmPart6726665
$107.41
Draeger Hydrogen Sulfide 0.2/a Detector Tubes, Box of 10
Draeger Hydrogen Sulfide 0.2/a Detector Tubes, Box of 10Handbook: 10 strokes, approx. 5 min · Sold by the box
Range0.2 to 5 ppmPart8101461
$113.65

Every Draeger tube we list, with the handbook data, is on the Draeger tubes A–Z chart; each gas has its own page with the OSHA and NIOSH limits.

Questions people ask

How many strokes does a Draeger tube need?

The number printed on the tube next to its scale; the handbook entry for every tube states it. Our tube pages show it under specifications.

How much air is one stroke?

100 mL (± 5 %) on the accuro and the X-act 5000 (handbook pp. 30, 33).

Can I use fewer strokes to save time?

The defined measuring range applies only with the stated strokes on a Draeger pump (p. 114). Where Draeger allows a different count, the entry states the extended range and the factor to apply.

How long does a Draeger tube reading take?

Between 10 s and 15 min for short-term tubes; each entry states the average time for its range (p. 35).

Why does Draeger say to use only a Draeger pump?

Because tube and pump are calibrated together as one system; the handbook states that mixing makers can impair the measurement (pp. 28–29).

Standards and ratings on this page are as published in the documents named and by the manufacturers named. The employer’s hazard assessment and written program decide what applies at a given workplace. Verify a rating on the product and its spec sheet before use.

Need a tube for a gas that is not listed, or a pump?Call (925) 215-7337 or email us with the gas and the range you expect. Orders ship in 1–2 business days.
All Draeger tubes A–Z
How to buyCard, purchase order or invoice terms; quotes on request; agencies and institutions welcome.
Live Service & SupportReal people who know the products, Monday to Friday 7:30am to 4pm Pacific. 925-215-7337
We DeliverWe ship nationwide. Parcel and LTL freight to job sites and businesses across the U.S.