Description
The Draeger 8103381 is the Acetone 40/a detector tube, measuring range 40 to 800 ppm, 1 pump strokes, measuring time approx. 1 min (handbook, p. 120). Sold by the box of 5, 5 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Acetone.
Draeger 8103381 specifications
| Item | Draeger data |
|---|---|
| Part number | 8103381 |
| Gas or substance | Acetone |
| Tube type | 40/a |
| Measuring range | 40 to 800 ppm |
| Number of strokes (n) | 1 |
| Measuring time | approx. 1 min |
| Standard deviation | ± 15 to 20 % |
| Color change | light yellow → yellow |
| Temperature | 5 to 40 °C |
| Humidity | 5 to 40 mg/L |
| Reaction principle | Acetone + 2,4-dinitrophenylhydrazine → yellow hydrazone |
| Pack | Box of 5 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 120, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 120): Other ketones are also indicated, but with different sensitivities. Aldehydes are also indicated. 500 ppm ethyl acetate does not affect the reading. Ammonia interferes with the indication due to a yellow-brown color- ing of the indicating layer. St-569-2008
Acetone exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 1000 ppm (2400 mg/m³) |
| NIOSH REL | TWA 250 ppm (590 mg/m³) |
| NIOSH IDLH | 2500 ppm [10%LEL] |
| CAS number | 67-64-1 |
| Formula | (CH₃)₂CO |
| Conversion | 1 ppm = 2.38 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Acetone (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Acetone
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| CH22901 | Draeger Acetone 100/b Detector Tubes, Box of 10 | 100 to 12000 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Acetone detector tubes.
Draeger 8103381 questions
What does the Draeger 8103381 measure? Acetone in air, measuring range 40 to 800 ppm.
How many pump strokes does the 8103381 take? 1 strokes (n) per the handbook, p. 120. Measuring time approx. 1 min.
Which pump does the 8103381 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the 8103381? light yellow → yellow, read against the scale printed on the tube.
Do Draeger 8103381 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of 8103381? 5. Each tube is used once.
Guides: what Draeger tubes measure · pump strokes and measuring time · how to read a tube · expiration and storage · cross-sensitivity · all tubes by gas and range.