Description
The Draeger CH22901 is the Acetone 100/b detector tube, measuring range 100 to 12000 ppm, 10 pump strokes, measuring time approx. 4 min (handbook, p. 121). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Acetone.
Draeger CH22901 specifications
| Item | Draeger data |
|---|---|
| Part number | CH22901 |
| Gas or substance | Acetone |
| Tube type | 100/b |
| Measuring range | 100 to 12000 ppm |
| Number of strokes (n) | 10 |
| Measuring time | approx. 4 min |
| Standard deviation | ± 15 to 20 % |
| Color change | light yellow → yellow |
| Temperature | 0 to 40 °C |
| Humidity | 5 to 20 mg H2O/L |
| Reaction principle | Acetone + 2,4-dinitrophenylhydrazine → yellow hydrazone |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 121, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 121): Other ketones are also indicated, but with different sensitivities. Aldehydes are also indicated, but not esters. Ammonia affects the reading, as it discolors the indicating layer yellow-brown.
Acetone exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 1000 ppm (2400 mg/m³) |
| NIOSH REL | TWA 250 ppm (590 mg/m³) |
| NIOSH IDLH | 2500 ppm [10%LEL] |
| CAS number | 67-64-1 |
| Formula | (CH₃)₂CO |
| Conversion | 1 ppm = 2.38 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Acetone (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Acetone
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| 8103381 | Draeger Acetone 40/a Detector Tubes, Box of 5 | 40 to 800 ppm | Box of 5 |
Every Draeger tube for this gas, with live prices: Acetone detector tubes.
Draeger CH22901 questions
What does the Draeger CH22901 measure? Acetone in air, measuring range 100 to 12000 ppm.
How many pump strokes does the CH22901 take? 10 strokes (n) per the handbook, p. 121. Measuring time approx. 4 min.
Which pump does the CH22901 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the CH22901? light yellow → yellow, read against the scale printed on the tube.
Do Draeger CH22901 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of CH22901? 10. Each tube is used once.