Description
The Draeger CH21201 is the Phosphine 50/a detector tube, measuring range 50 to 1000 ppm, 3 pump strokes, measuring time approx. 2 min (handbook, p. 252). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Phosphine.
Draeger CH21201 specifications
| Item | Draeger data |
|---|---|
| Part number | CH21201 |
| Gas or substance | Phosphine |
| Tube type | 50/a |
| Measuring range | 50 to 1000 ppm |
| Number of strokes (n) | 3 |
| Measuring time | approx. 2 min |
| Standard deviation | ± 10 to 15 % |
| Color change | yellow → brown-black |
| Temperature | 0 to 50 °C |
| Humidity | less than 40 mg H2O/L |
| Reaction principle | PH3 + Au3+ → Au (colloidal) |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 252, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 252): Arsine and stibine are also indicated, but with different sensitivities. Hydrogen sulfide, mercaptans, ammonia, and hydrochloric acid do not have any effect within their OEL ranges. Carbon monoxide and sulfur dioxide also have no effect in their OEL ranges.
Measuring range extension (handbook, p. 252): Measuring range 15 to 300 ppm where n = 10 strokes, multiply indicated scale value by 0.3.
Phosphine exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 0.3 ppm (0.4 mg/m³) |
| NIOSH REL | TWA 0.3 ppm (0.4 mg/m³) ST 1 ppm (1 mg/m³) |
| NIOSH IDLH | 50 ppm |
| CAS number | 7803-51-2 |
| Formula | PH₃ |
| Conversion | 1 ppm = 1.39 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Phosphine (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Phosphine
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| 8101611 | Draeger Phosphine 0.01/a Detector Tubes, Box of 10 | 0.1 to 1.0 ppm / 0.01 to 0.3 ppm | Box of 10 |
| 8103341 | Draeger Phosphine 0.1/b in Acetylene Detector Tubes, Box of 10 | 0.1 to 1 ppm / 1 to 15 ppm | Box of 10 |
| 8103711 | Draeger Phosphine 0.1/c Detector Tubes, Box of 10 | 0.5 to 3 ppm / 0.1 to 1.0 ppm | Box of 10 |
| 8101801 | Draeger Phosphine 1/a Detector Tubes, Box of 10 | 10 to 100 ppm / 1 to 20 ppm | Box of 10 |
| 8101621 | Draeger Phosphine 25/a Detector Tubes, Box of 10 | 200 to 10000 ppm / 25 to 900 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Phosphine detector tubes.
Draeger CH21201 questions
What does the Draeger CH21201 measure? Phosphine in air, measuring range 50 to 1000 ppm.
How many pump strokes does the CH21201 take? 3 strokes (n) per the handbook, p. 252. Measuring time approx. 2 min.
Which pump does the CH21201 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the CH21201? yellow → brown-black, read against the scale printed on the tube.
Do Draeger CH21201 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of CH21201? 10. Each tube is used once.