Description
The Draeger 8101721 is the Vinyl chloride 0.5/b detector tube, measuring range 5 to 30 ppm / 0.5 to 5 ppm, 1 /5 pump strokes, measuring time approx. 30 s / approx. 2.5 min (handbook, p. 274). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Vinyl chloride.
Draeger 8101721 specifications
| Item | Draeger data |
|---|---|
| Part number | 8101721 |
| Gas or substance | Vinyl chloride |
| Tube type | 0.5/b |
| Measuring range | 5 to 30 ppm / 0.5 to 5 ppm |
| Number of strokes (n) | 1 /5 |
| Measuring time | approx. 30 s / approx. 2.5 min |
| Standard deviation | ± 15 to 20 % |
| Color change | white → purple |
| Temperature | 10 to 30 °C |
| Humidity | max. 20 mg H2O/L |
| Reaction principle | a) CH2=CHCl + CrVI → Cl2 b) Cl2 + dimethylnaphtidin → purple reaction product |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 274, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 274): 100 ppm hydrogen chloride, 20 ppm chlorine, 10 ppm carbon tetrachloride, 10 ppm chloroform, or 5 ppm perchloroethylene are not indicated. Trichloroethylene and chlorobenzene are indicated with a lower sensitivity (5 ppm = reading of approx. 1.5 ppm). 1,1-dichloroethene is indicated with approximately the same sensi- tivity. Part of the oxidation layer is consumed under the influence of vapors of organic solvents, resulting in a correspondingly lower reading. Examples: 5 ppm vinyl chloride + 100 ppm butadiene or 5 ppm vinyl chloride + 10 ppm ethylene give a reading of 0.5 ppm vinyl chloride.
Vinyl chloride exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | [1910.1017] TWA 1 ppm C 5 ppm [15-minute] |
| NIOSH REL | Ca |
| NIOSH IDLH | Ca [N.D.] |
| CAS number | 75-01-4 |
| Formula | CH₂=CHCl |
| Conversion | 1 ppm = 2.56 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Vinyl chloride (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Vinyl chloride
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| CH19601 | Draeger Vinyl Chloride 100/a Detector Tubes, Box of 10 | 100 to 3000 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Vinyl chloride detector tubes.
Draeger 8101721 questions
What does the Draeger 8101721 measure? Vinyl chloride in air, measuring range 5 to 30 ppm / 0.5 to 5 ppm.
How many pump strokes does the 8101721 take? 1 /5 strokes (n) per the handbook, p. 274. Measuring time approx. 30 s / approx. 2.5 min.
Which pump does the 8101721 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the 8101721? white → purple, read against the scale printed on the tube.
Do Draeger 8101721 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of 8101721? 10. Each tube is used once.
Guides: what Draeger tubes measure · pump strokes and measuring time · how to read a tube · expiration and storage · cross-sensitivity · all tubes by gas and range.