Description
The Draeger CH19601 is the Vinyl chloride 100/a detector tube, measuring range 100 to 3000 ppm, 18 to 1 pump strokes, measuring time max. 3 min (handbook, p. 275). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Vinyl chloride.
Draeger CH19601 specifications
| Item | Draeger data |
|---|---|
| Part number | CH19601 |
| Gas or substance | Vinyl chloride |
| Tube type | 100/a |
| Measuring range | 100 to 3000 ppm |
| Number of strokes (n) | 18 to 1 |
| Measuring time | max. 3 min |
| Standard deviation | ± 30 % |
| Color change | purple → light brown |
| Temperature | 0 to 40 °C |
| Humidity | less than 30 mg H2O/L |
| Reaction principle | CH2=CHCl + MnO4– → MnIV + var. oxidation products |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 275, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 275): Many organic compounds with C=C double bonds are also indicat- ed, but all with different sensitivities. Differentiation is not possible. A vinyl chloride measurement is not possible in the presence of dialkyl sulfides.
Vinyl chloride exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | [1910.1017] TWA 1 ppm C 5 ppm [15-minute] |
| NIOSH REL | Ca |
| NIOSH IDLH | Ca [N.D.] |
| CAS number | 75-01-4 |
| Formula | CH₂=CHCl |
| Conversion | 1 ppm = 2.56 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Vinyl chloride (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Vinyl chloride
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| 8101721 | Draeger Vinyl Chloride 0.5/b Detector Tubes, Box of 10 | 5 to 30 ppm / 0.5 to 5 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Vinyl chloride detector tubes.
Draeger CH19601 questions
What does the Draeger CH19601 measure? Vinyl chloride in air, measuring range 100 to 3000 ppm.
How many pump strokes does the CH19601 take? 18 to 1 strokes (n) per the handbook, p. 275. Measuring time max. 3 min.
Which pump does the CH19601 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the CH19601? purple → light brown, read against the scale printed on the tube.
Do Draeger CH19601 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of CH19601? 10. Each tube is used once.
Guides: what Draeger tubes measure · pump strokes and measuring time · how to read a tube · expiration and storage · cross-sensitivity · all tubes by gas and range.