Description
The Draeger 8101811 is the Carbon dioxide 100/a detector tube, measuring range 100 to 3000 ppm, 10 pump strokes, measuring time approx. 4 min (handbook, p. 138). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Carbon dioxide.
Draeger 8101811 specifications
| Item | Draeger data |
|---|---|
| Part number | 8101811 |
| Gas or substance | Carbon dioxide |
| Tube type | 100/a |
| Measuring range | 100 to 3000 ppm |
| Number of strokes (n) | 10 |
| Measuring time | approx. 4 min |
| Standard deviation | ± 10 to 15 % |
| Color change | white to light |
| Temperature | 15 to 25 °C |
| Humidity | max. 23 mg H2O/L |
| Reaction principle | CO2 + N2H4 → NH2-NH-COOH |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 138, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 138): Hydrogen sulfide and sulfur dioxide do not provide an indication in their OEL ranges.
Carbon dioxide exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 5000 ppm (9000 mg/m³) |
| NIOSH REL | TWA 5000 ppm (9000 mg/m³) ST 30,000 ppm (54,000 mg/m³) |
| NIOSH IDLH | 40,000 ppm |
| CAS number | 124-38-9 |
| Formula | CO₂ |
| Conversion | 1 ppm = 1.80 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Carbon dioxide (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Carbon dioxide
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| CH23501 | Draeger Carbon Dioxide 0.1%/a Detector Tubes, Box of 10 | 0.5 to 6 Vol% / 0.1 to 1.2 Vol% | Box of 10 |
| CH31401 | Draeger Carbon Dioxide 0.5%/a Detector Tubes, Box of 10 | 0.5 to 10 Vol% | Box of 10 |
| CH25101 | Draeger Carbon Dioxide 1%/a Detector Tubes, Box of 10 | 1 to 20 Vol% | Box of 10 |
| 6728521 | Draeger Carbon Dioxide 100/a-P Tubes for Aerotest, Box of 10 | 100 to 3000 ppm | Box of 10 |
| CH20301 | Draeger Carbon Dioxide 5%/a Detector Tubes, Box of 10 | 5 to 60 Vol% | Box of 10 |
Every Draeger tube for this gas, with live prices: Carbon dioxide detector tubes.
Draeger 8101811 questions
What does the Draeger 8101811 measure? Carbon dioxide in air, measuring range 100 to 3000 ppm.
How many pump strokes does the 8101811 take? 10 strokes (n) per the handbook, p. 138. Measuring time approx. 4 min.
Which pump does the 8101811 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the 8101811? white to light, read against the scale printed on the tube.
Do Draeger 8101811 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of 8101811? 10. Each tube is used once.