Description
The Draeger CH23501 is the Carbon dioxide 0.1%/a detector tube, measuring range 0.5 to 6 Vol% / 0.1 to 1.2 Vol%, 1 /5 pump strokes, measuring time approx. 30 s / approx. 2.5 min (handbook, p. 139). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Carbon dioxide.
Draeger CH23501 specifications
| Item | Draeger data |
|---|---|
| Part number | CH23501 |
| Gas or substance | Carbon dioxide |
| Tube type | 0.1%/a |
| Measuring range | 0.5 to 6 Vol% / 0.1 to 1.2 Vol% |
| Number of strokes (n) | 1 /5 |
| Measuring time | approx. 30 s / approx. 2.5 min |
| Standard deviation | ± 5 to 10 % |
| Color change | white → purple |
| Temperature | 0 to 30 °C |
| Humidity | max. 30 mg H2O/L |
| Reaction principle | CO2 + amine → purple reaction product |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 139, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 139): Hydrogen sulfide and sulfur dioxide do not provide an indication in their OEL ranges.
Carbon dioxide exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 5000 ppm (9000 mg/m³) |
| NIOSH REL | TWA 5000 ppm (9000 mg/m³) ST 30,000 ppm (54,000 mg/m³) |
| NIOSH IDLH | 40,000 ppm |
| CAS number | 124-38-9 |
| Formula | CO₂ |
| Conversion | 1 ppm = 1.80 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Carbon dioxide (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Carbon dioxide
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| CH31401 | Draeger Carbon Dioxide 0.5%/a Detector Tubes, Box of 10 | 0.5 to 10 Vol% | Box of 10 |
| CH25101 | Draeger Carbon Dioxide 1%/a Detector Tubes, Box of 10 | 1 to 20 Vol% | Box of 10 |
| 8101811 | Draeger Carbon Dioxide 100/a Detector Tubes, Box of 10 | 100 to 3000 ppm | Box of 10 |
| 6728521 | Draeger Carbon Dioxide 100/a-P Tubes for Aerotest, Box of 10 | 100 to 3000 ppm | Box of 10 |
| CH20301 | Draeger Carbon Dioxide 5%/a Detector Tubes, Box of 10 | 5 to 60 Vol% | Box of 10 |
Every Draeger tube for this gas, with live prices: Carbon dioxide detector tubes.
Draeger CH23501 questions
What does the Draeger CH23501 measure? Carbon dioxide in air, measuring range 0.5 to 6 Vol% / 0.1 to 1.2 Vol%.
How many pump strokes does the CH23501 take? 1 /5 strokes (n) per the handbook, p. 139. Measuring time approx. 30 s / approx. 2.5 min.
Which pump does the CH23501 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the CH23501? white → purple, read against the scale printed on the tube.
Do Draeger CH23501 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of CH23501? 10. Each tube is used once.
Guides: what Draeger tubes measure · pump strokes and measuring time · how to read a tube · expiration and storage · cross-sensitivity · all tubes by gas and range.