Description
The Draeger CH24201 is the Sulfur dioxide 20/a detector tube, measuring range 20 to 200 ppm, 10 pump strokes, measuring time approx. 3 min (handbook, p. 260). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Sulfur dioxide.
Draeger CH24201 specifications
| Item | Draeger data |
|---|---|
| Part number | CH24201 |
| Gas or substance | Sulfur dioxide |
| Tube type | 20/a |
| Measuring range | 20 to 200 ppm |
| Number of strokes (n) | 10 |
| Measuring time | approx. 3 min |
| Standard deviation | ± 10 to 15 % |
| Color change | brown-yellow → white |
| Temperature | 0 to 40 °C |
| Humidity | less than 30 mg H2O/L |
| Reaction principle | SO2 + I2 + 2 H2O → H2SO4 + 2 HI |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 260, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 260): A sulfur dioxide measurement is not possible in the presence of hydrogen sulfide, as hydrogen sulfide is indicated with roughly the same sensitivity. Nitrogen dioxide shortens the reading.
Measuring range extension (handbook, p. 260): Measuring range 200 to 2000 ppm where n = 1 stroke, multiply the indicated value by 10. During the 1-stroke measurement, 3 desorption strokes in air free of sulfur dioxide need to be performed afterwards.
Sulfur dioxide exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 5 ppm (13 mg/m³) |
| NIOSH REL | TWA 2 ppm (5 mg/m³) ST 5 ppm (13 mg/m³) |
| NIOSH IDLH | 100 ppm |
| CAS number | 7446-09-5 |
| Formula | SO₂ |
| Conversion | 1 ppm = 2.62 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Sulfur dioxide (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Sulfur dioxide
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| 6727101 | Draeger Sulfur Dioxide 0.1/a Detector Tubes, Box of 10 | 0.1 to 3 ppm | Box of 10 |
| 6728491 | Draeger Sulfur Dioxide 0.5/a Detector Tubes, Box of 10 | 1 to 25 ppm / 0.5 to 5 ppm | Box of 10 |
| CH31701 | Draeger Sulfur Dioxide 1/a Detector Tubes, Box of 10 | 1 to 25 ppm | Box of 10 |
| 8101531 | Draeger Sulfur Dioxide 50/b Detector Tubes, Box of 10 | 400 to 8000 ppm / 50 to 500 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Sulfur dioxide detector tubes.
Draeger CH24201 questions
What does the Draeger CH24201 measure? Sulfur dioxide in air, measuring range 20 to 200 ppm.
How many pump strokes does the CH24201 take? 10 strokes (n) per the handbook, p. 260. Measuring time approx. 3 min.
Which pump does the CH24201 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the CH24201? brown-yellow → white, read against the scale printed on the tube.
Do Draeger CH24201 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of CH24201? 10. Each tube is used once.
Guides: what Draeger tubes measure · pump strokes and measuring time · how to read a tube · expiration and storage · cross-sensitivity · all tubes by gas and range.