Description
The Draeger CH29101 is the Hydrogen sulfide 100/a detector tube, measuring range 100 to 2000 ppm, 1 pump strokes, measuring time approx. 30 s (handbook, p. 210). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Hydrogen sulfide.
Draeger CH29101 specifications
| Item | Draeger data |
|---|---|
| Part number | CH29101 |
| Gas or substance | Hydrogen sulfide |
| Tube type | 100/a |
| Measuring range | 100 to 2000 ppm |
| Number of strokes (n) | 1 |
| Measuring time | approx. 30 s |
| Standard deviation | ± 5 to 10 % |
| Color change | white → brown |
| Temperature | 0 to 40 °C |
| Humidity | 3 to 40 mg H2O/L |
| Reaction principle | H2S + Pb2+ → PbS + 2 H+ |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 210, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 210): 2000 ppm sulfur dioxide and 100 ppm nitrogen dioxide have no effect.
Hydrogen sulfide exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | C 20 ppm 50 ppm [10-minute maximum peak] |
| NIOSH REL | C 10 ppm (15 mg/m³) [10-minute] |
| NIOSH IDLH | 100 ppm |
| CAS number | 7783-06-4 |
| Formula | H₂S |
| Conversion | 1 ppm = 1.40 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Hydrogen sulfide (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Hydrogen sulfide
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| CH28101 | Draeger Hydrogen Sulfide 0.2%/A Detector Tubes, Box of 10 | 0.2 to 7 Vol% | Box of 10 |
| 8101461 | Draeger Hydrogen Sulfide 0.2/a Detector Tubes, Box of 10 | 0.2 to 5 ppm | Box of 10 |
| 8101991 | Draeger Hydrogen Sulfide 0.2/b Detector Tubes, Box of 10 | 0.2 to 6 ppm | Box of 10 |
| 6728041 | Draeger Hydrogen Sulfide 0.5/a Detector Tubes, Box of 10 | 0.5 to 15 ppm / 5 to 150 ppm | Box of 10 |
| 6719001 | Draeger Hydrogen Sulfide 1/c Detector Tubes, Box of 10 | 10 to 200 ppm / 1 to 20 ppm | Box of 10 |
| 8101831 | Draeger Hydrogen Sulfide 1/d Detector Tubes, Box of 10 | 10 to 200 ppm / 1 to 20 ppm | Box of 10 |
| 8101211 | Draeger Hydrogen Sulfide 2%/a Detector Tubes, Box of 10 | 2 to 40 Vol% | Box of 10 |
| 6728821 | Draeger Hydrogen Sulfide 2/a Detector Tubes, Box of 10 | 20 to 200 ppm / 2 to 20 ppm | Box of 10 |
| 8101961 | Draeger Hydrogen Sulfide 2/b Detector Tubes, Box of 10 | 2 to 60 ppm | Box of 10 |
| CH29801 | Draeger Hydrogen Sulfide 5/b Detector Tubes, Box of 10 | 5 to 60 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Hydrogen sulfide detector tubes.
Draeger CH29101 questions
What does the Draeger CH29101 measure? Hydrogen sulfide in air, measuring range 100 to 2000 ppm.
How many pump strokes does the CH29101 take? 1 strokes (n) per the handbook, p. 210. Measuring time approx. 30 s.
Which pump does the CH29101 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the CH29101? white → brown, read against the scale printed on the tube.
Do Draeger CH29101 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of CH29101? 10. Each tube is used once.