Description
The Draeger CH29901 is the Carbon monoxide 0.3%/b detector tube, measuring range 0.3 to 7 Vol%, 1 pump strokes, measuring time approx. 30 s (handbook, p. 150). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Carbon monoxide.
Draeger CH29901 specifications
| Item | Draeger data |
|---|---|
| Part number | CH29901 |
| Gas or substance | Carbon monoxide |
| Tube type | 0.3%/b |
| Measuring range | 0.3 to 7 Vol% |
| Number of strokes (n) | 1 |
| Measuring time | approx. 30 s |
| Standard deviation | ± 10 to 15 % |
| Color change | white → brown-green |
| Temperature | 0 to 50 °C |
| Humidity | max. 50 mg H2O/L |
| Reaction principle | H2S2O7 5 CO + I2O5 → I2 + 5 CO2 |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 150, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 150): The following have no effect on the reading of 0.3 Vol% CO: 10000 ppm n-octane, 300 ppm benzene, 500 ppm hydrogen sulfide, 500 ppm sulfur dioxide, 500 ppm nitrogen dioxide, 300 ppm butadiene, 250 ppm chloroform, 3000 ppm acetylene give a reading of 0.3 Vol%.
Carbon monoxide exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 50 ppm (55 mg/m³) |
| NIOSH REL | TWA 35 ppm (40 mg/m³) C 200 ppm (229 mg/m³) |
| NIOSH IDLH | 1200 ppm |
| CAS number | 630-08-0 |
| Formula | CO |
| Conversion | 1 ppm = 1.15 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Carbon monoxide (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Carbon monoxide
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| CH20601 | Draeger Carbon Monoxide 10/b Detector Tubes, Box of 10 | 100 to 3000 ppm / 10 to 300 ppm | Box of 10 |
| 6733051 | Draeger Carbon Monoxide 2/a Detector Tubes, Box of 10 | 2 to 60 ppm / 25 to 300 ppm | Box of 10 |
| 3735950 | Draeger Carbon Monoxide 2/a-P Tubes for Aerotest, Box of 10 | 2 to 60 ppm | Box of 10 |
| 6728511 | Draeger Carbon Monoxide 5/a-P Tubes for Aerotest, Box of 10 | 5-150 ppm | Box of 10 |
| CH25601 | Draeger Carbon Monoxide 5/c Detector Tubes, Box of 10 | 100 to 700 ppm / 5 to 150 ppm | Box of 10 |
| CH19701 | Draeger Carbon Monoxide 8/a Detector Tubes, Box of 10 | 8 to 150 ppm CO in H2 | Box of 10 |
Every Draeger tube for this gas, with live prices: Carbon monoxide detector tubes.
Draeger CH29901 questions
What does the Draeger CH29901 measure? Carbon monoxide in air, measuring range 0.3 to 7 Vol%.
How many pump strokes does the CH29901 take? 1 strokes (n) per the handbook, p. 150. Measuring time approx. 30 s.
Which pump does the CH29901 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the CH29901? white → brown-green, read against the scale printed on the tube.
Do Draeger CH29901 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of CH29901? 10. Each tube is used once.