Description
The Draeger 6728071 is the Benzene 5/b detector tube, measuring range 5 to 50 ppm, 20 pump strokes, measuring time approx. 8 min (handbook, p. 136). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Benzene.
Draeger 6728071 specifications
| Item | Draeger data |
|---|---|
| Part number | 6728071 |
| Gas or substance | Benzene |
| Tube type | 5/b |
| Measuring range | 5 to 50 ppm |
| Number of strokes (n) | 20 |
| Measuring time | approx. 8 min |
| Standard deviation | ± 10 to 15 % |
| Color change | white → brown-green |
| Temperature | 0 to 40 °C |
| Humidity | 3 to 15 mg H2O/L |
| Reaction principle | C6H6 + I2O5 → I2 |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 136, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 136): Many gasoline hydrocarbons are also indicated, but all with different sensitivities. Differentiation is not possible. Other aromatics are also indicated.
Benzene exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | [1910.1028] TWA 1 ppm ST 5 ppm |
| NIOSH REL | Ca TWA 0.1 ppm ST 1 ppm |
| NIOSH IDLH | Ca [500 ppm] |
| CAS number | 71-43-2 |
| Formula | C₆H₆ |
| Conversion | 1 ppm = 3.19 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Benzene (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Benzene
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| 8103511 | Draeger 8103511 PID Pre-Filter Tubes for Benzene, Box of 10 | Pre-filter tube that lets a PID read benzene selectively (not a measuring tube) | Box of 10 |
| 8103691 | Draeger Benzene 0.25/a Detector Tubes, Box of 10 | 0.25 to 2 ppm / 2 to 10 ppm | Box of 10 |
| 8101741 | Draeger Benzene 15/a Detector Tubes, Box of 10 | 15 to 420 ppm | Box of 10 |
| 8101231 | Draeger Benzene 2/a Detector Tubes, Box of 5 | 2 to 60 ppm | Box of 5 |
| 6718801 | Draeger Benzene 5/a Detector Tubes, Box of 10 | 5 to 40 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Benzene detector tubes.
Draeger 6728071 questions
What does the Draeger 6728071 measure? Benzene in air, measuring range 5 to 50 ppm.
How many pump strokes does the 6728071 take? 20 strokes (n) per the handbook, p. 136. Measuring time approx. 8 min.
Which pump does the 6728071 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the 6728071? white → brown-green, read against the scale printed on the tube.
Do Draeger 6728071 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of 6728071? 10. Each tube is used once.
Guides: what Draeger tubes measure · pump strokes and measuring time · how to read a tube · expiration and storage · cross-sensitivity · all tubes by gas and range.
