Description
The Draeger 8101021 is the Carbon tetrachloride 1/a detector tube, measuring range 1 to 15 ppm / 10 to 50 ppm, 10 /5 pump strokes, measuring time approx. 6 min / 3 min (handbook, p. 152). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Carbon tetrachloride.
Draeger 8101021 specifications
| Item | Draeger data |
|---|---|
| Part number | 8101021 |
| Gas or substance | Carbon tetrachloride |
| Tube type | 1/a |
| Measuring range | 1 to 15 ppm / 10 to 50 ppm |
| Number of strokes (n) | 10 /5 |
| Measuring time | approx. 6 min / 3 min |
| Standard deviation | ± 15 to 20 % |
| Color change | white → yellow |
| Temperature | 15 to 30 °C |
| Humidity | 3 to 15 mg H2O/L |
| Reaction principle | a) CCl4 + H2S2O7 → COCl2 b) COCl2 + arom. nitro compound → yellow reaction product |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 152, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 152): Chloropicrin and phosgene are indicated with roughly the same sensitivity, so a carbon tetrachloride measurement is not possible in their presence. The indication is not affected by: 1 ppm chlorine 5 ppm hydrochloric acid 20 ppm methyl bromide 1000 ppm acetone
Carbon tetrachloride exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 10 ppm C 25 ppm 200 ppm (5-minute maximum peak in any 4 hours) |
| NIOSH REL | Ca ST 2 ppm (12.6 mg/m³) [60-minute] |
| NIOSH IDLH | Ca [200 ppm] |
| CAS number | 56-23-5 |
| Formula | CCl₄ |
| Conversion | 1 ppm = 6.29 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Carbon tetrachloride (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Carbon tetrachloride
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| 8103501 | Draeger Carbon Tetrachloride 0.1/a Detector Tubes, Box of 10 | 0.1 to 5 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Carbon tetrachloride detector tubes.
Draeger 8101021 questions
What does the Draeger 8101021 measure? Carbon tetrachloride in air, measuring range 1 to 15 ppm / 10 to 50 ppm.
How many pump strokes does the 8101021 take? 10 /5 strokes (n) per the handbook, p. 152. Measuring time approx. 6 min / 3 min.
Which pump does the 8101021 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the 8101021? white → yellow, read against the scale printed on the tube.
Do Draeger 8101021 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of 8101021? 10. Each tube is used once.
Guides: what Draeger tubes measure · pump strokes and measuring time · how to read a tube · expiration and storage · cross-sensitivity · all tubes by gas and range.