Description
The Draeger 8101891 is the Carbon disulfide 3/a detector tube, measuring range 3 to 95 ppm, 15 to 1 pump strokes, measuring time max. 2 min (handbook, p. 143). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Carbon disulfide.
Draeger 8101891 specifications
| Item | Draeger data |
|---|---|
| Part number | 8101891 |
| Gas or substance | Carbon disulfide |
| Tube type | 3/a |
| Measuring range | 3 to 95 ppm |
| Number of strokes (n) | 15 to 1 |
| Measuring time | max. 2 min |
| Standard deviation | ± 30 % |
| Color change | light blue → yellow-green |
| Temperature | 0 to 40 °C |
| Humidity | less than 30 mg H2O/L |
| Reaction principle | 2 CS2 + 4 NHR2 + Cu2+ → Cu(SCSNR2)2 + 2 NH2R2+ |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 143, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 143): Hydrogen sulfide is retained in the pre-layer at concentrations around the OEL, so has no effect.
Carbon disulfide exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 20 ppm C 30 ppm 100 ppm (30-minute maximum peak) |
| NIOSH REL | TWA 1 ppm (3 mg/m³) ST 10 ppm (30 mg/m³) [skin] |
| NIOSH IDLH | 500 ppm |
| CAS number | 75-15-0 |
| Formula | CS₂ |
| Conversion | 1 ppm = 3.11 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Carbon disulfide (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Carbon disulfide
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| CH23201 | Draeger Carbon Disulfide 30/a Detector Tubes, Box of 10 | 0.1 to 10 mg/L | Box of 10 |
| 6728351 | Draeger Carbon Disulfide 5/a Detector Tubes, Box of 10 | 5 to 60 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Carbon disulfide detector tubes.
Draeger 8101891 questions
What does the Draeger 8101891 measure? Carbon disulfide in air, measuring range 3 to 95 ppm.
How many pump strokes does the 8101891 take? 15 to 1 strokes (n) per the handbook, p. 143. Measuring time max. 2 min.
Which pump does the 8101891 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the 8101891? light blue → yellow-green, read against the scale printed on the tube.
Do Draeger 8101891 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of 8101891? 10. Each tube is used once.