Description
The Draeger CH23201 is the Carbon disulfide 30/a detector tube, measuring range 0.1 to 10 mg/L, 6 pump strokes, measuring time approx. 1 min (handbook, p. 145). Sold by the box of 10, 10 single-use tubes. Read on a Draeger accuro or X-act 5000 tube pump. Ships in 1–2 business days; most orders arrive in 3–7 business days. All Draeger tubes for Carbon disulfide.
Draeger CH23201 specifications
| Item | Draeger data |
|---|---|
| Part number | CH23201 |
| Gas or substance | Carbon disulfide |
| Tube type | 30/a |
| Measuring range | 0.1 to 10 mg/L |
| Number of strokes (n) | 6 |
| Measuring time | approx. 1 min |
| Standard deviation | ± 15 to 20 % |
| Color change | light blue → brown |
| Temperature | 0 to 40 °C |
| Humidity | < 30 mg H2O/L |
| Reaction principle | 2 CS2 + 4 NHR2 + Cu2+ → Cu(SCSNR2)2 + 2 NH2R2+ |
| Pack | Box of 10 |
| Pump | Read on a Draeger accuro or X-act 5000 tube pump |
Source: Dräger-Tubes & MicroTubes Handbook, 22nd edition (order no. 90 92 086), p. 145, Dräger Safety AG & Co. KGaA.
Cross-sensitivity (handbook, p. 145): A carbon disulfide measurement is not possible in the presence of hydrogen sulfide, as hydrogen sulfide colors the indicating layer light green.
Carbon disulfide exposure limits
| Limit | Value |
|---|---|
| OSHA PEL | TWA 20 ppm C 30 ppm 100 ppm (30-minute maximum peak) |
| NIOSH REL | TWA 1 ppm (3 mg/m³) ST 10 ppm (30 mg/m³) [skin] |
| NIOSH IDLH | 500 ppm |
| CAS number | 75-15-0 |
| Formula | CS₂ |
| Conversion | 1 ppm = 3.11 mg/m³ |
Source: NIOSH Pocket Guide to Chemical Hazards, Carbon disulfide (OSHA PEL from 29 CFR 1910.1000 Table Z as listed there). TWA = 8-hour time-weighted average, ST = short-term (15-minute), C = ceiling, Ca = NIOSH potential occupational carcinogen.
Other Draeger tubes for Carbon disulfide
| Part number | Tube | Measuring range | Pack |
|---|---|---|---|
| 8101891 | Draeger Carbon Disulfide 3/a Detector Tubes, Box of 10 | 3 to 95 ppm | Box of 10 |
| 6728351 | Draeger Carbon Disulfide 5/a Detector Tubes, Box of 10 | 5 to 60 ppm | Box of 10 |
Every Draeger tube for this gas, with live prices: Carbon disulfide detector tubes.
Draeger CH23201 questions
What does the Draeger CH23201 measure? Carbon disulfide in air, measuring range 0.1 to 10 mg/L.
How many pump strokes does the CH23201 take? 6 strokes (n) per the handbook, p. 145. Measuring time approx. 1 min.
Which pump does the CH23201 use? A Draeger tube pump: the accuro hand pump or the X-act 5000. Per the handbook (p. 28–29), Draeger short-term tubes and Draeger tube pumps are matched at the factory and form a single unit; using other pumps with Draeger tubes, or other tubes with Draeger pumps, can impair the measurement.
What is the color change on the CH23201? light blue → brown, read against the scale printed on the tube.
Do Draeger CH23201 tubes expire? Yes. Draeger prints an expiration date (month and year) on the package; the handbook states a typical shelf life for the customer of up to two years, stored sealed at the temperature on the package strip (handbook section 2.13).
How many tubes are in a box of CH23201? 10. Each tube is used once.